Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8021B: CAL ADJ /5100 INSTRUMENT: Fluke 8022B: CAL ADJ /5100 DATE: 10-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 63 CONFIGURATION: Fluke 5100B ============================================================================= # # Source: # Fluke 8021B instruction manual (PN 616037 July 1981). # Fluke 8022B instruction manual (PN 616045 July 1981). # # Compatibility: # MET/CAL 3.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J F W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP Ambient temperature: 21C - 25C. 1.003 DISP Relative humidity: less than 80%. 1.003 DISP Allow the UUT to stabilize for 30 minutes. 1.004 HEAD CALIBRATION ACCESS PROCEDURE 1.005 DISP Disconnect the test leads and the battery eliminator 1.005 DISP if attached. Open the battery compartment, and 1.005 DISP disconnect the battery. Remove the three phillips-head 1.005 DISP screws from the bottom of the case. Turn the instrument 1.005 DISP face-up, and grasp the top cover at both sides of the 1.005 DISP input connectors. Then, pull the top cover from the UUT. 1.006 DISP Re-connect the battery and turn the UUT on. 1.007 HEAD DC VOLTAGE CALIBRATION ADJUSTMENT 1.008 DISP Connect the 5100B and UUT as follows: 1.008 DISP [32] 5100B OUTPUT HI to UUT V Ohm 1.008 DISP [32] 5100B OUTPUT LO to UUT COMMON 1.009 RNG 200mV 1.010 5100 190mV S 2W 1.011 DISP Adjust R5 (DC CAL) for a display of 190.0 or 190.1. 1.012 5100 * S 1.013 HEAD AC VOLTAGE CALIBRATION ADJUSTMENT 1.014 RNG 200mV AC 1.015 5100 190mV 100H S 2W 1.016 DISP Adjust R9 (AC CAL) for a display of 190.0 1.016 DISP (an occasional flash of +/-1 digit is acceptable). 1.017 END